RecallRadar

FDA Device recall Z-2251-2020

Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate…

On 2020-06-10 the FDA classified a Class II recall of Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate… by Abbott Laboratories, citing in the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample di….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2251-2020
ClassificationClass II
Statusterminated
Initiated2020-04-02
Published2020-06-10
Last updated2026-09-13 11:55 UTC
CompanyAbbott Laboratories
DistributionUS

Product

Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.

Reason

In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.

Identifiers

code_infoAll serial numbers in the field

Status history