FDA Device recall Z-2251-2012

OEC IT3000

On 2012-09-05 the FDA classified a Class I recall of OEC IT3000 by Ge Oec Medical Systems, citing GE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, an….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2251-2012
ClassificationClass I
Statusterminated
Initiated2009-06-22
Published2012-09-05
Last updated2026-09-13 11:53 UTC
CompanyGe Oec Medical Systems
DistributionUS

Product

OEC IT3000; IT2500 (EnTrak); IT2500 Plus (EnTrak Plus); IT3500 (InstaTrak); IT3500 Plus (InstaTrak Plus); ConneCTstat; and ConneCTstat Plus picture archiving and communications systems. The systems are an aid to locate anatomical structures during open or percutaneous surgical procedures. It is intended for use in medical conditions that may benefit from the use of stereotactic surgical technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.

Reason

GE Healthcare Surgery had discovered that using the Inverted Headset Placement is not a validated configuration with the IT 3000, 2500, 2500 plus, 3500, 3500 plus, ConneCTstat, and ConneCTstat Plus Surgical Navigation equipment.

Identifiers

code_infoAll serial numbers.