RecallRadar

FDA Device recall Z-2244-2021

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811)

On 2021-08-25 the FDA classified a Class I recall of Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811) by Cardinal Health 200, citing the current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2244-2021
ClassificationClass I
Statusterminated
Initiated2021-06-16
Published2021-08-25
Last updated2026-09-13 11:56 UTC
CompanyCardinal Health 200
DistributionUS

Product

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Reason

The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

Identifiers

code_infoAll lot codes

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.