FDA Device recall Z-2240-2023

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pedia…

On 2023-08-02 the FDA classified a Class II recall of Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pedia… by Siemens Medical Solutions Usa, citing incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2240-2023
ClassificationClass II
Statusongoing
Initiated2023-06-05
Published2023-08-02
Last updated2026-09-13 11:56 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088

Reason

Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).

Identifiers

code_infoUDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2240-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.