FDA Device recall Z-2238-2026
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA
- FDA Device
- Class I
- ongoing
- Published 2026-06-03
On 2026-06-03 the FDA classified a Class I recall of Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA by Abiomed, citing retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in cer….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2238-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2011-01-17 |
| Published | 2026-06-03 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Abiomed |
| Distribution | US |
Product
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
Reason
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Identifiers
| code_info | 1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC…1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715. 2. Product Code: 0042-0000-EU. GTIN: 00813502011289. Serial Numbers: IC1217, IC1265, IC1275, IC1276, IC1288, IC1284, IC1301, IC1302, IC1304, IC1305, IC1344, IC1341, IC1343, IC1340, IC1345, IC1350, IC1347, IC1349, IC1383, IC1381, IC1384, IC1386, IC1393, IC1396, IC1397, IC1395, IC1413, IC1415, IC1411, IC1416, IC1490, IC1488, IC1522, IC1523, IC1524, IC1525, IC1547, IC1549, IC1546, IC1548, IC1578, IC1577, IC1579, IC1580, IC1594, IC1584, IC1590, IC1592, IC1591, IC1585, IC1588, IC1653, IC1676, IC1685, IC1684, IC1677, IC1678, IC1699, IC1700. 3. Product Code: 0042-0000-UK. GTIN: 00813502011296. Serial Numbers: IC1387, IC1389, IC1388, IC1519, IC1542, IC1545, IC1581, IC1582, IC1583, IC1573. 4. Product Code: 0042-0000-US. GTIN: 00813502010022. Serial Numbers: IC1004, IC1002, IC1001, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2238-2026%22