On 2020-06-10 the FDA classified a Class II recall of Beckman Coulter Access Immunoassay Systems, Access Free T3 Triiodothyronine, REF A13422 (UDI: 15099590201661) for use with the Access Famil… by Beckman Coulter, citing the recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A1342….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2238-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-04-30 |
| Published | 2020-06-10 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Beckman Coulter |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
Product
Beckman Coulter Access Immunoassay Systems, Access Free T3 Triiodothyronine, REF A13422 (UDI: 15099590201661) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.
Reason
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
Identifiers
| code_info | Catalog Number: A13422; all lots. |
|---|
Status history