FDA Device recall Z-2238-2018
G-PIN 2.8BL 3.2PL 98CM 2PK
- FDA Device
- Class II
- terminated
- Published 2018-06-27
On 2018-06-27 the FDA classified a Class II recall of G-PIN 2.8BL 3.2PL 98CM 2PK by Zimmer Biomet, citing the firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier fai….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2238-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-05-16 |
| Published | 2018-06-27 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zimmer Biomet |
| Distribution | US |
Product
G-PIN 2.8BL 3.2PL 98CM 2PK; 467260 NL DRIVING GUIDE 4.4MM X 80CM; 467220 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Reason
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
Identifiers
| code_info | 143206 281765 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2238-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.