RecallRadar

FDA Device recall Z-2237-2020

Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with Access Family of Immunoas…

On 2020-06-10 the FDA classified a Class II recall of Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with Access Family of Immunoas… by Beckman Coulter, citing the recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A1342….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2237-2020
ClassificationClass II
Statusterminated
Initiated2019-04-30
Published2020-06-10
Last updated2026-09-13 11:55 UTC
CompanyBeckman Coulter
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

Reason

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

Identifiers

code_infoCatalog Number: 33880; all lots.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2237-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.