FDA Device recall Z-2233-2012
Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the sy…
- FDA Device
- Class II
- terminated
- Published 2012-08-29
On 2012-08-29 the FDA classified a Class II recall of Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the sy… by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terum….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2233-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-07-26 |
| Published | 2012-08-29 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Interface module for CDl System 500 Product Usage: The Interface Module for CDl 500 is the interface between the CDl 500 monitor and the system. The module performs the following functions: Communicates primary pump flow data to the CDl 500 Receives blood parameter data from the CDl 500
Reason
Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
Identifiers
| code_info | catalog number: 803479 and serial numbers:00009-00122, 00124-00210, 00213-00365, 00367-00458, and 00461-00596. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2233-2012%22