RecallRadar

FDA Device recall Z-2231-2023

Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy Singl…

On 2023-08-02 the FDA classified a Class II recall of Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy Singl… by Boston Scientific, citing manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Conti….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2231-2023
ClassificationClass II
Statusongoing
Initiated2023-05-25
Published2023-08-02
Last updated2026-09-13 11:56 UTC
CompanyBoston Scientific
DistributionUS

Product

Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy Single-pack UPN: M006L8406910

Reason

Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched

Identifiers

code_infoGTIN: 08714729978671 Lot Numbers: 30389427, 30389429, 30389431, 30521475, 30521476, 30521477 Exp. Date: 20-Oct-2025 to 10-Nov-2025

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2231-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.