FDA Device recall Z-2230-2021
Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used thr…
- FDA Device
- Class II
- completed
- Published 2021-08-18
On 2021-08-18 the FDA classified a Class II recall of Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used thr… by Shent Usa, citing mislabeled: Devices labeled as catalog #7208 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm) may actually be catalog #7226 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2230-2021 |
| Classification | Class II |
| Status | completed |
| Initiated | 2021-01-12 |
| Published | 2021-08-18 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Shent Usa |
| Distribution | US |
Product
Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used through the working channel of a GI endoscope
Reason
Mislabeled: Devices labeled as catalog #7208 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm) may actually be catalog #7226 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 8 mm x 230 cm) and Devices labeled as catalog #7226, (Grasper/Retriever, Alligator Jaw, Rat Tooth, 2.4 mm x 8 mm x 230 cm) may actually be catalog #7208, Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm).
Identifiers
| code_info | Lot Code: D20177226-19 |
|---|
Status history