FDA Device recall Z-2230-2012
Serial interface module RS-232 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module
- FDA Device
- Class II
- terminated
- Published 2012-08-29
On 2012-08-29 the FDA classified a Class II recall of Serial interface module RS-232 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terum….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2230-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-07-26 |
| Published | 2012-08-29 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
Serial interface module RS-232 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module. This module is a general purpose serial interface and is the interface between the TLink Data Management ystem or Sarns CCPro Data Perfusion oftware and the system. The data collected and exported by the data transfer module includes physiological data and device events to record the progress and events during a procedure.
Reason
Terumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
Identifiers
| code_info | catalog number: 802113 and serial numbers: 00007-00036, 00038-00045, 00047-00064, 00066-00077, 00080-00121, 00123-00214…catalog number: 802113 and serial numbers: 00007-00036, 00038-00045, 00047-00064, 00066-00077, 00080-00121, 00123-00214,00216-00283, 00285-00363, 00366-00498, 00500-00548, and 00550-00597. |
|---|