FDA Device recall Z-2227-2012

Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in…

On 2012-08-29 the FDA classified a Class III recall of Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in… by Beckman Coulter, citing A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to sam….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2227-2012
ClassificationClass III
Statusterminated
Initiated2011-10-11
Published2012-08-29
Last updated2026-09-13 11:53 UTC
CompanyBeckman Coulter
DistributionUS

Product

Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in human serum and plasma in the assessment of thyroid function and diagnosis of thyroid disorders when T3 levels are between 0.1 and 8.0 ng/mL.

Reason

A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.

Identifiers

code_infoAll Lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2227-2012%22