FDA Device recall Z-2223-2012
The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body…
- FDA Device
- Class II
- terminated
- Published 2012-08-29
On 2012-08-29 the FDA classified a Class II recall of The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body… by Philips Medical Systems Cleveland, citing 1. The current system design of software version 4.0.0xx379 does not prevent the system from scanning if the dose level exceeds a set upper dose limit for CT Brain Perfusion scann….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2223-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-07-18 |
| Published | 2012-08-29 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Philips Medical Systems Cleveland |
| Distribution | IN, OH |
Product
The Brilliance 64 and Ingenuity CT are whole body Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Reason
1. The current system design of software version 4.0.0xx379 does not prevent the system from scanning if the dose level exceeds a set upper dose limit for CT Brain Perfusion scanning, which previous versions did. However, the current version does have the Dose Check (NEMA XR-25) feature, which provides dose notifications and alerts, but it is possible for the user to disable Dose Check after ente
Identifiers
| code_info | System Code # 728231, Serial #s: 91003, 95157, 9654, 95130, and 95504 System Code # 728326, Serial #s: 300010, 300030, and 300018 |
|---|