RecallRadar

FDA Device recall Z-2221-2025

Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P

On 2025-08-06 the FDA classified a Class II recall of Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P by Volcano, citing reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may requir….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2221-2025
ClassificationClass II
Statusongoing
Initiated2025-06-13
Published2025-08-06
Last updated2026-09-13 11:57 UTC
CompanyVolcano
DistributionUS

Product

Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

Reason

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Identifiers

code_infoUDI-DI: 00845225002848, 00184360000037. Instructions for use: 300004875732_A Revision Date: 03/2021, 300004410212.A Release Date: 03/2021

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2221-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.