FDA Device recall Z-2218-2012
System 8000 Base Product Usage: The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circu…
- FDA Device
- Class III
- terminated
- Published 2012-08-22
On 2012-08-22 the FDA classified a Class III recall of System 8000 Base Product Usage: The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circu… by Terumo Cardiovascular Systems, citing during a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 wit….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2218-2012 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2012-06-20 |
| Published | 2012-08-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | US |
Product
System 8000 Base Product Usage: The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Reason
During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating identified on the name plate label. The mislabeling was identified in January 2007 a set up process of the automark label printer. All units were corrected in the field in Marc
Identifiers
| code_info | catalog number: 16401 and serial number: 2001, 2002, and 2003. |
|---|