RecallRadar

FDA Device recall Z-2217-2023

SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Descript…

On 2023-08-02 the FDA classified a Class II recall of SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Descript… by Datascope, citing during IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2217-2023
ClassificationClass II
Statusongoing
Initiated2023-06-23
Published2023-08-02
Last updated2026-09-13 11:56 UTC
CompanyDatascope
DistributionUS

Product

SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0469-01C SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0469-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0470-01C SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0470-07 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0470-09 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES D684-00-0470-01 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0470-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0469-01U SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES D684-00-0469-01 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY D684-00-0469-09 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs D884-00-0019-16

Reason

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Identifiers

code_info
Product Description UDI SENSATION…Product Description UDI SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (CHINA) 10607567109367 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES (INDIA) 10607567109732 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (CHINA) 10607567109374 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (INDIA) 10607567109749 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES (TURKEY) 10607567113135 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES 10607567106779 SENSATION 7 Fr. 40cc IAB WITH ACCESSORIES, US ONLY 10607567109558 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, US ONLY 10607567109541 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES 10607567106755 SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES, TURKEY 10607567113128 PACKAGED INSERTION KIT - SENSATION 7 Fr. 34/40 cc IABs 10607567106724

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2217-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.