RecallRadar

FDA Device recall Z-2217-2016

Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Dia…

On 2016-07-27 the FDA classified a Class II recall of Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Dia… by Carestream Health, citing A software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2217-2016
ClassificationClass II
Statusterminated
Initiated2016-07-08
Published2016-07-27
Last updated2026-09-13 11:54 UTC
CompanyCarestream Health
DistributionGA, IA, TX

Product

Carestream Touch Prime, Catalog Number 1738830, and Carestream Touch Prime XE, Catalog Number 1738822 --- Common Name: Touch Ultrasound Diagnostic ultrasound imaging or fluid flow analysis of the human body

Reason

A software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of the anatomy may appear enlarged or distended when it actually is not.

Identifiers

code_infoTouch Prime Serial Numbers: 1037, 1038; -- Touch Prime XE Serial Numbers: 1021, 1079, 1067, 1041, 1043

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2217-2016%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.