RecallRadar

FDA Device recall Z-2215-2012

00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete…

On 2012-08-22 the FDA classified a Class II recall of 00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete… by Zimmer, citing lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors m….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2215-2012
ClassificationClass II
Statusterminated
Initiated2012-07-18
Published2012-08-22
Last updated2026-09-13 11:53 UTC
CompanyZimmer
DistributionUS

Product

00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65¿¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."

Reason

Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain

Identifiers

code_info
Part 00-4309-029-01: Lot 60535162, 60545338, 60576488, 60605855, 60637022, 60689253, 60772974, 60875277, 60918060,…Part 00-4309-029-01: Lot 60535162, 60545338, 60576488, 60605855, 60637022, 60689253, 60772974, 60875277, 60918060, 60989817, 61011689, 61024161, and 61051024.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2215-2012%22