FDA Device recall Z-2213-2014
Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is intended as an intraoperative image-guided localization system to enable minimally inva…
- FDA Device
- Class II
- terminated
- Published 2014-08-20
On 2014-08-20 the FDA classified a Class II recall of Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is intended as an intraoperative image-guided localization system to enable minimally inva… by Brainlab Ag, citing brainlab Navigation Software Spine & Trauma 3D 2.0/2.1 offers automatic registration of intraoperatively acquired CT image data sets. To enable automatic registration, the sof….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2213-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-05-21 |
| Published | 2014-08-20 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Brainlab Ag |
| Distribution | US |
Product
Navigation Spine & Trauma 3D Version 2.0 and 2.1 Is intended as an intraoperative image-guided localization system to enable minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or Intraoperative 2D or 3D image data. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy.
Reason
Brainlab Navigation Software Spine & Trauma 3D 2.0/2.1 offers automatic registration of intraoperatively acquired CT image data sets. To enable automatic registration, the software requires the gantry position of the scanner. The gantry position can either be entered manually or submitted automatically from the CT scanner. If automatic gantry communication is available, the so-called "
Identifiers
| code_info | Model/Catalog Number: 19061-03 |
|---|
Status history