FDA Device recall Z-2212-2026
Pediatric care bed
- FDA Device
- Class I
- ongoing
- Published 2026-06-10
On 2026-06-10 the FDA classified a Class I recall of Pediatric care bed by Kayserbetten And Kg, citing if the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased ri….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2212-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-05-04 |
| Published | 2026-06-10 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Kayserbetten And Kg |
| Distribution | US |
Product
Pediatric care bed; Product Designation: KayserBett IDA;
Reason
If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).
Identifiers
| code_info | Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636…Product Designation: KayserBett IDA; UDI-DI: 426038961IDAGU; Serial Number: 386, 387, 388, 389, 390, 391, 392, 635, 636, 637, 638, 639, 641, 642, 643, 644, 645, 646, 647, 1072, 1073, 1074, 1075, 1076, 1077, 1078, 1079, 1080, 1081, 1082, 1083, 1085, 1086, 1087, 1088, 1089, 1090, 1110, 1111, 1446, 1447, 1448, 1449, 1648, 1649, 1950, 1975, 1976, 1982, 1983, 1998, 2001, 2003, 2004, 2337, 2338, 2339, 2340, 2341, 2342, 2343, 2344, 2345, 2396, 2763, 2764, 2765, 2766, 2767, 2770, 2771, 2772, 2773, 2774, 3087, 3088, 3089, 3090, 3091, 3092, 3167, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3309, 3310, 3311, 3312, 3313, 3314, 3740, 3741, 3760, 4223, 4232, 4233, 4234, 4235, 4236, 4237, 4239, 4240, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4986, 5238, 5254, 5255, 5256, 5257, 5258, 5259, 5637, 5919, 5920, 5921, 5922, 5923, 5924, 5925, 5926, 5937, 5938, 5939, 5940, 5948, 6332, 6 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2212-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.