FDA Device recall Z-2211-2014
Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219
- FDA Device
- Class II
- terminated
- Published 2014-08-20
On 2014-08-20 the FDA classified a Class II recall of Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219 by Medical Positioning, citing excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a c….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2211-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-07-10 |
| Published | 2014-08-20 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Medical Positioning |
| Distribution | US |
Product
Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations.
Reason
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
Identifiers
| code_info | Serial number--803976 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2211-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.