RecallRadar

FDA Device recall Z-2210-2016

Preamendments device-Sarns Dual Cooler-Heater

On 2016-07-27 the FDA classified a Class II recall of Preamendments device-Sarns Dual Cooler-Heater by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Produc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2210-2016
ClassificationClass II
Statusterminated
Initiated2016-06-08
Published2016-07-27
Last updated2026-09-13 11:54 UTC
CompanyTerumo Cardiovascular Systems
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY

Product

Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for supplying cold and warm water to heat exchangers during open heart surgery. It supplies water at a rate of up to 6.5 gallons per minute (25 liters per minute) at zero head, and will provide a maximum pressure output of 11 pounds per square inch (76,000 pascals). The Sarns TM Dual Cooler-Heater can be used with heat exchangers capable of accepting these flows and pressures. Product Usage: The Sarns Dual Cooler-Heater is indicated for use as a water supply system to a blood heat exchanger in open-heart surgery.

Reason

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Identifiers

code_info
The Sarns Dual Cooler-Heater ; Dual Cooler Heater 115V/60Hz (P/N 11160); Dual Cooler Heater 220V/50Hz (P/N 4332); Dua…The Sarns Dual Cooler-Heater ; Dual Cooler Heater 115V/60Hz (P/N 11160); Dual Cooler Heater 220V/50Hz (P/N 4332); Dual Cooler Heater 1 OOV/60Hz (P/N 14114); Dual Cooler Heater 1 OOV/50Hz (PIN 13460)