FDA Device recall Z-2208-2014

Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017

On 2014-08-20 the FDA classified a Class II recall of Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017 by Medical Positioning, citing excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a c….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2208-2014
ClassificationClass II
Statusterminated
Initiated2014-07-10
Published2014-08-20
Last updated2026-09-13 11:53 UTC
CompanyMedical Positioning
DistributionUS

Product

Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage: Used to conduct syncope examinations.

Reason

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

Identifiers

code_info
Serial number--802971, 803106, 803356, 803588, 803616, 803854, 802781, 802783, 802780, 802784, 802421, 802861, 802939,…Serial number--802971, 803106, 803356, 803588, 803616, 803854, 802781, 802783, 802780, 802784, 802421, 802861, 802939, 802949, 802924, 803867, 803926, 802809 and 804118.