FDA Device recall Z-2207-2016
HX2" Temperature Management Systems
- FDA Device
- Class II
- terminated
- Published 2016-07-27
On 2016-07-27 the FDA classified a Class II recall of HX2" Temperature Management Systems by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2207-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-08 |
| Published | 2016-07-27 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Terumo Cardiovascular Systems |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY |
Product
HX2" Temperature Management Systems; The Terumo¿ HX2" Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right).The system has the capacity to circulate water at a rate of up to 6.5 gal.min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Product Usage: HX2 Temperature Management Systems is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.
Reason
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
Identifiers
| code_info | HX2" Temperature Management Systems, P/N 809810, Catalog Number 809810 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2207-2016%22