FDA Device recall Z-2207-2015

ADVIA Centaur Tni-Ultra Assay

On 2015-07-29 the FDA classified a Class II recall of ADVIA Centaur Tni-Ultra Assay by Siemens Healthcare Diagnostics, citing system-to-system bias between the Tni-Ultra assay on the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2207-2015
ClassificationClass II
Statusterminated
Initiated2015-05-19
Published2015-07-29
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

ADVIA Centaur Tni-Ultra Assay; SMN 10317708 - 100 tests and SMN 10317709 -500 tests Product Usage: For in vitro diagnostic use in the quantitative determination of cardiac troponin I (cTnl) in serum, heparinized plasma, or EDTA plasma using the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT/ADVIA Centaur CP systems. Cardiac troponin I determinations aid in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction (MI) or increased probability of ischemic events requiring urgent revascularization procedures.

Reason

System-to-system bias between the Tni-Ultra assay on the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems.

Identifiers

code_info
***This field action applies to all in-date lots of reagents (Kit lots ending in 088, 089, 090, 091, 093 and 094) and a…***This field action applies to all in-date lots of reagents (Kit lots ending in 088, 089, 090, 091, 093 and 094) and all future lots until the issue is resolved and a follow-up communication is issued by Siemens.**** SMN 10317708 - 100 tests 43796099 May 25, 2015 43818088 May 25, 2015 44793088 May 25, 2015 45687089 June 22, 2015 46527089 June 22, 2015 48096089 June 22, 2015 49216090 August3, 2015 50779090 August3, 2015 52113091 September 14, 2015 53091093 October 19, 2015 54684093 October 19, 2015 SMN 10317709 -500 tests 43797088 May25, 2015 43819088 May25,2015 44098088 May25, 2015 44623088 May 25, 2015 45688089 June 22, 2015 46661089 June 22, 2015 48301090 August3, 2015 49869090 August 3, 2015 50884091 September 14, 2015 51471091 September 14, 2015 51798091 September 14, 2015 53092093 October 19,2015 53231093 October 19,2015 54742094 December 6, 2015

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.