RecallRadar

FDA Device recall Z-2206-2026

ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600

On 2026-06-10 the FDA classified a Class II recall of ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600 by Siemens Medical Solutions Usa, citing during patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2206-2026
ClassificationClass II
Statusongoing
Initiated2026-03-09
Published2026-06-10
Last updated2026-09-13 11:58 UTC
CompanySiemens Medical Solutions Usa
Distributionn/a

Product

ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system

Reason

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Identifiers

code_infoARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2206-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.