FDA Device recall Z-2202-2020
AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is i…
- FDA Device
- Class II
- terminated
- Published 2020-06-10
On 2020-06-10 the FDA classified a Class II recall of AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is i… by Agfa N V, citing under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2202-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-04-06 |
| Published | 2020-06-10 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Agfa N V |
| Distribution | US |
Product
AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications
Reason
Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible
Identifiers
| code_info | Serial Numbers: 0925, 0945, 0948, 0949, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, and 1404 |
|---|