RecallRadar

FDA Device recall Z-2200-2026

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-C…

On 2026-05-20 the FDA classified a Class II recall of LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-C… by Microvention, citing stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2200-2026
ClassificationClass II
Statusongoing
Initiated2026-04-06
Published2026-05-20
Last updated2026-09-13 11:58 UTC
CompanyMicrovention
Distributionn/a

Product

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm

Reason

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

Identifiers

code_info
Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)0…Lots: 0000616462, 0000619851, 0000552987, 0000551725. UDI: (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115893(11)240325(17)270228(10)0000552987, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240325(17)270228(10)0000551725, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115930(11)240610(17)270531(10)0000616462, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851, (01)00842429115961(11)240610(17)270531(10)0000619851,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2200-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.