FDA Device recall Z-2200-2020
Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instru…
- FDA Device
- Class II
- ongoing
- Published 2020-06-10
On 2020-06-10 the FDA classified a Class II recall of Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instru… by Inpeco, citing in certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2200-2020 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-04-07 |
| Published | 2020-06-10 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Inpeco |
| Distribution | US |
Product
Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.
Reason
In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.
Identifiers
| code_info | Serial Numbers: AP2.0005 AP2.0008 AP2.0010 AP2.0014 AP2.0017 AP2.0018 AP2.0022 AP2.0023 AP2.0026 AP2.0027 AP…Serial Numbers: AP2.0005 AP2.0008 AP2.0010 AP2.0014 AP2.0017 AP2.0018 AP2.0022 AP2.0023 AP2.0026 AP2.0027 AP2.0033 AP2.0036 AP2.0039 AP2.0043 AP2.0045 AP2.0056 AP2.0060 AP2.0065 AP2.0069 AP2.0070 AP2.0071 AP2.0076 AP2.0078 AP2.0081 AP2.0087 AP2.0089 AP2.0102 AP2.0104 AP2.0105 AP2.0110 AP2.0111 AP2.0113 AP2.0120 AP2.0127 AP2.0137 AP2.0139 AP2.0146 AP2.0147 AP2.0149 AP2.0151 AP2.0155 AP2.0165 AP2.0170 AP2.0172 AP2.0184 AP2.0189 AP2.0201 AP2.0202 AP2.0203 AP2.0209 AP2.0211 AP2.0213 AP2.0217 AP2.0223 APT.0009 APT.0014 APT.0058 APT.0065 APT.0093 APT.0091 APT.0125 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2200-2020%22