FDA Device recall Z-2191-2012
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 131HF7P. For use in patients who require hemodynamic monitoring
- FDA Device
- Class II
- terminated
- Published 2012-08-22
On 2012-08-22 the FDA classified a Class II recall of Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 131HF7P. For use in patients who require hemodynamic monitoring by Edwards Lifesciences, citing the device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2191-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-05-11 |
| Published | 2012-08-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Edwards Lifesciences |
| Distribution | US |
Product
Edwards Lifesciences Swan-Ganz Thermodilution VIP Catheters Model 131HF7P. For use in patients who require hemodynamic monitoring.
Reason
The device is being recalled because the firm received multiple complaints related to passage of a J-tip guidewire when inserted through the distal lumen hub of the catheter.
Identifiers
| code_info | 59169527, 59169528, 59174318, 59174320, 59194314, 59194315, 59194316, 59194317, 59197004, 59197005, 59197006, 59197007,…59169527, 59169528, 59174318, 59174320, 59194314, 59194315, 59194316, 59194317, 59197004, 59197005, 59197006, 59197007, 59197008, 59197009, 59197010, 59197897, 59197898, 59197899, 59197900, 59197901, 59197902, 59197903, 59203374. 59203375, 59203377, 59203378, 59203379, 59203380, 59203381, 59207074, 59207075, 59207076, 59207077, 59216092, 59216093, 59216094, 59216096, 59216097, 59216098, 59216099, 59219846, 59230687, 59230688, 59230689, 59230690, 59230691, 59230692, 59233882, 59238962, 59238963, 59238964, 59238965, 59243250, 59243249, 59247419, 59247421, 59247420, 59247422, 59247423, 59247424, 59247425, 59247426, 59247427, 59247428, 59247429 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2191-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.