FDA Device recall Z-2187-2021
REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866
- FDA Device
- Class II
- ongoing
- Published 2021-08-11
On 2021-08-11 the FDA classified a Class II recall of REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866 by Abbott Laboratories, citing there is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in pat….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2187-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-06-10 |
| Published | 2021-08-11 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Abbott Laboratories |
| Distribution | US |
Product
REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866
Reason
There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results
Identifiers
| code_info | Lot Numbers / Expiration Date 84126UN20 / 09JUL2021 03676UN20 / 25AUG2021 10597UN20 / 29AUG2021 26275UN20 / 20OCT…Lot Numbers / Expiration Date 84126UN20 / 09JUL2021 03676UN20 / 25AUG2021 10597UN20 / 29AUG2021 26275UN20 / 20OCT2021 77748UN20 / 08DEC2021 65958UN20 / 08DEC2021 81824UN21 / 16FEB2022 |
|---|
Status history