FDA Device recall Z-2186-2018

Accu-Chek Performa Test Strip, Blood glucose monitoring system, Catalog no. 07299702001 Product Usage: The Accu-Chek Performa test strips a…

On 2018-06-20 the FDA classified a Class II recall of Accu-Chek Performa Test Strip, Blood glucose monitoring system, Catalog no. 07299702001 Product Usage: The Accu-Chek Performa test strips a… by Roche Diabetes Care, citing the recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2186-2018
ClassificationClass II
Statusterminated
Initiated2018-05-07
Published2018-06-20
Last updated2026-09-13 11:54 UTC
CompanyRoche Diabetes Care
DistributionUS

Product

Accu-Chek Performa Test Strip, Blood glucose monitoring system, Catalog no. 07299702001 Product Usage: The Accu-Chek Performa test strips are for use with the Accu-Chek Performa meter to quantitatively measure glucose (sugar) in venous whole blood or fresh capillary whole blood samples from the fingertips.

Reason

The recalling firm has identified two lots of test strips that are out of specification. Affected test strips show an increased potential for: strip errors prior to applying a blood sample; strips not detected once inserted; or a biased result may occur, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result as compared to the actual blood glucose value.

Identifiers

code_infoLot 476605

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2186-2018%22