RecallRadar

FDA Device recall Z-2185-2021

REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161859

On 2021-08-11 the FDA classified a Class II recall of REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161859 by Abbott Laboratories, citing there is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in pat….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2185-2021
ClassificationClass II
Statusongoing
Initiated2021-06-10
Published2021-08-11
Last updated2026-09-13 11:56 UTC
CompanyAbbott Laboratories
DistributionUS

Product

REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161859

Reason

There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

Identifiers

code_infoLot Numbers / Expiration Date 02954UN20 / 29AUG2021 37921UN20 / 01DEC2021 81823UN21 / 16FEB2022

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2185-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.