RecallRadar

FDA Device recall Z-2185-2018

Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Us…

On 2018-06-20 the FDA classified a Class II recall of Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Us… by Roche Diabetes Care, citing the recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detec….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2185-2018
ClassificationClass II
Statusterminated
Initiated2018-05-07
Published2018-06-20
Last updated2026-09-13 11:54 UTC
CompanyRoche Diabetes Care
DistributionUS

Product

Accu-Chek Aviva Plus Test Strip, Blood glucose monitoring system, Catalog nos. 06908217001, 06908268001, 06908349001,06908217001 Product Usage: The Accu-Chek Aviva Plus test strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

Reason

The recalling firm identified four US lots of test strips out of specification. Affected strips show increased potential for strip errors prior to applying blood, strips not detected once inserted, or biased results, which may not be detectable. In the case of a biased result, incorrect action could be taken by the user in response to the displayed blood glucose result.

Identifiers

code_infoLots 496947, 497291, 497296, 497325

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.