FDA Device recall Z-2184-2021
REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, I…
- FDA Device
- Class II
- ongoing
- Published 2021-08-11
On 2021-08-11 the FDA classified a Class II recall of REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, I… by Abbott Laboratories, citing there is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in pati….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2184-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-06-10 |
| Published | 2021-08-11 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Abbott Laboratories |
| Distribution | US |
Product
REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569
Reason
There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results.
Identifiers
| code_info | Lot Numbers / Expiration Date 83459UN20 /09JUL2021, 03168UN20 /25SEP2021, 09787UN20 /24OCT2021, 37977UN20 /01JAN2…Lot Numbers / Expiration Date 83459UN20 /09JUL2021, 03168UN20 /25SEP2021, 09787UN20 /24OCT2021, 37977UN20 /01JAN2022, 63000UN20 /01JAN2022, 77745UN20 /08JAN2022, 77743UN21 /02MAR2022, 09237UN21 /09APR2022 |
|---|