FDA Device recall Z-2179-2014
Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian RE…
- FDA Device
- Class II
- terminated
- Published 2014-08-20
On 2014-08-20 the FDA classified a Class II recall of Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian RE… by Medtronic Minimed, citing medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2179-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-07-21 |
| Published | 2014-08-20 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Medtronic Minimed |
| Distribution | US |
Product
Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combination with the Minilink transmitter and Sof-sensor glucose sensor for continuous monitoring of subcutaneous glucose concentration. The information provided by these devices is intended for use in identifying potential episodes of high or low blood glucose. The information provided is not intended to be used directly for making therapy adjustment but rather to assist the user in determining that an additional fingerstick glucose measurement may be appropriate. Per the product labeling, all therapy adjustments should be based on fingerstick glucose measurements obtained using standard home glucose meters and not on the sensor glucose values provided by the Guardian REAL-Time monitor. The CSS7100 and CSS7100K monitors are identical expect for the fact that the CSS7100K monitor is a pediatric version that has a minimum programmable low alert setting of 90 mg/dL. The minimum programmable low glucose alert setting for the adult model (CSS7100) is 40 mg/dL.
Reason
Medtronic MiniMed is recalling the Guardian REAL-Time monitors because of a manufacturing issue that has the potential to compromise the water resistance of the affected Guardian monitors.
Identifiers
| code_info | CSS7100 and CSS7100K |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2179-2014%22