FDA Device recall Z-2174-2014

Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option

On 2014-08-20 the FDA classified a Class II recall of Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option by Siemens Medical Solutions Usa, citing A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-se….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2174-2014
ClassificationClass II
Statusterminated
Initiated2014-05-28
Published2014-08-20
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Siemens Linear Accelerators of type ARTISTE, ONCOR, and PRIMUS with Automatic Sequenced Cancer Treatment Delivery Option.

Reason

A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision with the patient.

Identifiers

code_info
material numbers: 01940035, 04504200, 05672977, 05693908, 05857912, 05857920, 08139789, 08319758, 08490005, 085…material numbers: 01940035, 04504200, 05672977, 05693908, 05857912, 05857920, 08139789, 08319758, 08490005, 08515520, 09401654, 09401746, 09822693, with serial numbers below 3094.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2174-2014%22