FDA Device recall Z-2174-2012
Shiley Reusable Cannula Low Pressure Cuffed Tracheostomy Tubes, model numbers (Worldwide) 8LPC, 8FEN
- FDA Device
- Class I
- terminated
- Published 2012-08-22
On 2012-08-22 the FDA classified a Class I recall of Shiley Reusable Cannula Low Pressure Cuffed Tracheostomy Tubes, model numbers (Worldwide) 8LPC, 8FEN by Nellcor Puritan Bennett Dba Covidien, citing covidien is recalling certain cuffed Shiley tracheostomy tubes due to consumer complaints related to the connections and functions of the inner and outer cannula which were observ….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2174-2012 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2012-07-18 |
| Published | 2012-08-22 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Nellcor Puritan Bennett Dba Covidien |
| Distribution | US |
Product
Shiley Reusable Cannula Low Pressure Cuffed Tracheostomy Tubes, model numbers (Worldwide) 8LPC, 8FEN; (Outside U.S.) 8FENJ, 8FENJ-S, 8LPC-S, 8FEN-S. Product Usage: The product is intended for use to bypass upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions. The product is sold as sterile. The packaging may indicate either Covidien or Tyco as the manufacturer as the firm has transitioned from Tyco to Covidien over the past 4 years.
Reason
Covidien is recalling certain cuffed Shiley tracheostomy tubes due to consumer complaints related to the connections and functions of the inner and outer cannula which were observed during mechanical ventilation of patients.
Identifiers
| code_info | United States distributed product includes: 8FEN lots 0910000342 to 1101001558 and 110300622X to 120600700X; 8LPC lots…United States distributed product includes: 8FEN lots 0910000342 to 1101001558 and 110300622X to 120600700X; 8LPC lots 0910000348 to 1101001823 and 110200327X to 120600351X. Product Distributed outside of the United States: 8FENJ lots 1003002176, 1004000412, and 120100433X. 8FEN-S lot 0910001010; 8FENJ-S lots 0910002075 to 1103002299 and 110601803X to 120402125X; and 8LPC-S lots 0910001002 to 1103002052 and 111000400X to 120600014X. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2174-2012%22