RecallRadar

FDA Device recall Z-2173-2024

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analy…

On 2024-07-03 the FDA classified a Class II recall of Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analy… by Sight Diagnostics, citing one Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results i….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2173-2024
ClassificationClass II
Statusongoing
Initiated2024-05-13
Published2024-07-03
Last updated2026-09-13 11:57 UTC
CompanySight Diagnostics
DistributionUS

Product

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Reason

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

Identifiers

code_infoREF: TK1, UDI-DI: B787TK10, Lot Code: 10114