FDA Device recall Z-2172-2018
Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria
- FDA Device
- Class II
- terminated
- Published 2018-06-20
On 2018-06-20 the FDA classified a Class II recall of Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria by Biomerieux, citing A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2172-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-03-23 |
| Published | 2018-06-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Biomerieux |
| Distribution | n/a |
Product
Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria. This enzyme is characteristic of the genus Neisseria and most species of Pseudomonas. It enables the differentiation of Gram (-) bacilli (1, 2). A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
Reason
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Identifiers
| code_info | Oxidase Reagent (ref: 55635): Lot number 92366601 |
|---|
Status history