FDA Device recall Z-2167-2025
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMI…
- FDA Device
- Class I
- ongoing
- Published 2025-08-06
On 2025-08-06 the FDA classified a Class I recall of Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMI… by Hamilton Medical Ag, citing ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2167-2025 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-06-26 |
| Published | 2025-08-06 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Hamilton Medical Ag |
| Distribution | US |
Product
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Reason
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.
Identifiers
| code_info | UDI-DI: 07630002802963, Lots: 199675, 199676 |
|---|