Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMI…

On 2025-08-06 the FDA classified a Class I recall of Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMI… by Hamilton Medical Ag, citing ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2167-2025
ClassificationClass I
Statusongoing
Initiated2025-06-26
Published2025-08-06
Last updated2026-09-13 11:57 UTC
CompanyHamilton Medical Ag
DistributionUS

Product

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Reason

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.

Identifiers

code_infoUDI-DI: 07630002802963, Lots: 199675, 199676

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.