FDA Device recall Z-2167-2017

Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a ran…

On 2017-06-07 the FDA classified a Class III recall of Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a ran… by Randox Laboratories, citing according to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2167-2017
ClassificationClass III
Statusterminated
Initiated2017-04-27
Published2017-06-07
Last updated2026-09-13 11:54 UTC
CompanyRandox Laboratories
DistributionPR

Product

Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.

Reason

According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice

Identifiers

code_infoCatalogue number UC5075 Batch number 907UC

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.