FDA Device recall Z-2167-2017
Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a ran…
- FDA Device
- Class III
- terminated
- Published 2017-06-07
On 2017-06-07 the FDA classified a Class III recall of Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a ran… by Randox Laboratories, citing according to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2167-2017 |
| Classification | Class III |
| Status | terminated |
| Initiated | 2017-04-27 |
| Published | 2017-06-07 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Randox Laboratories |
| Distribution | PR |
Product
Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.
Reason
According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice
Identifiers
| code_info | Catalogue number UC5075 Batch number 907UC |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.