FDA Device recall Z-2166-2018
The VITEK¿ 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK¿ 2 Systems for the automated identif…
- FDA Device
- Class II
- terminated
- Published 2018-06-20
On 2018-06-20 the FDA classified a Class II recall of The VITEK¿ 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK¿ 2 Systems for the automated identif… by Biomerieux, citing A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2166-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-03-23 |
| Published | 2018-06-20 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Biomerieux |
| Distribution | n/a |
Product
The VITEK¿ 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK¿ 2 Systems for the automated identification of most clinically significant anaerobic organisms and Corynebatcerium species. The VITEK¿ 2 ANC identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
Reason
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Identifiers
| code_info | VITEK¿ 2 ANC (ref: 21347): Lot Number 2440318203 |
|---|