RecallRadar

FDA Device recall Z-2164-2026

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software V…

On 2026-05-20 the FDA classified a Class II recall of Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software V… by Dexcom, citing A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the ap….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2164-2026
ClassificationClass II
Statusongoing
Initiated2026-04-14
Published2026-05-20
Last updated2026-09-13 11:58 UTC
CompanyDexcom
DistributionUS

Product

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used are the same for each user. The watchOS app cannot be installed independently from the iOS phone app. Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System

Reason

A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.

Identifiers

code_infoLot Code: SW13355 versions 2.9.0 through 2.11.2 - UDI-DI: 00386270005168

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2164-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.