FDA Device recall Z-2163-2017

CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the s…

On 2017-05-31 the FDA classified a Class II recall of CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the s… by Corpak Medsystems, citing firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2163-2017
ClassificationClass II
Statusterminated
Initiated2017-02-03
Published2017-05-31
Last updated2026-09-13 11:54 UTC
CompanyCorpak Medsystems
DistributionCA, NH, VT

Product

CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.

Reason

Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.

Identifiers

code_infoCatalogue Numbers: 50-6016