FDA Device recall Z-2160-2026
RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U…
- FDA Device
- Class I
- ongoing
- Published 2026-06-03
On 2026-06-03 the FDA classified a Class I recall of RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U… by Bolton Medical, citing labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2160-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-04-22 |
| Published | 2026-06-03 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Bolton Medical |
| Distribution | PR |
Product
RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-154-34U 28-N4-34-209-30U 28-N4-34-209-34U 28-N4-34-259-34U 28-N4-36-109-36U 28-N4-36-154-32U 28-N4-36-154-36U 28-N4-36-199-32U 28-N4-36-199-36U 28-N4-36-259-36U 28-N4-38-109-38U 28-N4-38-154-34U 28-N4-38-154-38U 28-N4-38-199-34U 28-N4-38-199-38U 28-N4-38-259-38U 28-N4-40-114-40U 28-N4-40-154-36U 28-N4-40-154-40U 28-N4-40-204-36U 28-N4-40-204-40U 28-N4-40-259-40U 28-N4-42-114-42U 28-N4-42-159-38U 28-N4-42-159-42U 28-N4-42-204-38U 28-N4-42-204-42U 28-N4-42-259-42U 28-N4-44-114-44U 28-N4-44-164-40U 28-N4-44-164-44U 28-N4-44-209-40U 28-N4-44-209-44U 28-N4-44-259-40U 28-N4-44-259-44U 28-N4-46-114-46U 28-N4-46-164-42U 28-N4-46-164-46U 28-N4-46-209-42U 28-N4-46-209-46U 28-N4-46-259-42U
Reason
Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.
Identifiers
| code_info | All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)008435761…All Lots/(Ref Number/Proximal Diameter/DI)28-N4-32-104-32U/ 32/ (01)00843576150461; 28-N4-32-164-28U/ 32/ (01)00843576150959; 28-N4-32-164-32U/ 32/ (01)00843576150591; 28-N4-32-209-28U/ 32/ (01)00843576151055; 28-N4-32-209-32U/ 32/ (01)00843576150720; 28-N4-32-259-32U/ 32/ (01)00843576150850; 28-N4-34-109-34U/ 34/ (01)00843576150478; 28-N4-34-154-30U/ 34/ (01)00843576150966; 28-N4-34-154-34U/ 34/ (01)00843576150607; 28-N4-34-209-30U/ 34/ (01)00843576151062; 28-N4-34-209-34U/ 34/ (01)00843576150737; 28-N4-34-259-34U/ 34/ (01)00843576150867; 28-N4-36-109-36U/ 36/ (01)00843576150485; 28-N4-36-154-32U/ 36/ (01)00843576150973; 28-N4-36-154-36U/ 36/ (01)00843576150614; 28-N4-36-199-32U/ 36/ (01)00843576151079; 28-N4-36-199-36U/ 36/ (01)00843576150744; 28-N4-36-259-36U/ 36/ (01)00843576150874; 28-N4-38-109-38U/ 38/ (01)00843576150492; 28-N4-38-154-34U/ 38/ (01)00843576150980; 28-N4-38-154-38U/ 38/ (01)00843576150621; 28-N4-38-199-34U/ 38/ (01)00843576151086; 28-N4-38-199-38U/ 38/ (01)00843576 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2160-2026%22