FDA Device recall Z-2159-2026
NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX,…
- FDA Device
- Class I
- ongoing
- Published 2026-05-20
On 2026-05-20 the FDA classified a Class I recall of NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX,… by Arrow International, citing 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2159-2026 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2026-04-10 |
| Published | 2026-05-20 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Arrow International |
| Distribution | US |
Product
NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-SFXM, CS-15232-SFXM; NextStep Retrograde Hemodialysis Catheter, REF: AC-15192-X, AC-15232-X, AC-15272-X, CS-15192-X, CS-15232-X, CS-15272-X, CS-15312-X; NextStep Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-XM, CS-15232-XM, CS-15272-XM, CS-15312-XM
Reason
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
Identifiers
| code_info | REF/UDI-DI/Lots: AC-15192-SFX/30801902197758/AC-15192-SFX; AC-15232-SFX/00801902095909/33F23H1018; CS-15192-SFX/308019…REF/UDI-DI/Lots: AC-15192-SFX/30801902197758/AC-15192-SFX; AC-15232-SFX/00801902095909/33F23H1018; CS-15192-SFX/30801902195297/33F23E0575, 33F23G0530, 33F23K0187, 33F23K1104, 33F23L0929, 33F24A0275, 33F24C0068, 33F24D1012, 33F25C0020, 33F25E1083, 33F25G0636, 33F25H0292, 33F25H0790, 33F25L0038; CS-15232-SFX/30801902195549/33F23E0576, 33F23G0534, 33F23L0936, 33F24A0279, 33F24C0067, 33F24D1006, 33F24E0684, 33F24H0258, 33F24J0033, 33F24J0085, 33F24J0342, 33F24L0056, 33F24M0231, 33F25A0405, 33F25C0036, 33F25C0664, 33F25D0739, 33F25E0433, 33F25G0716, 33F25H0248; CS-15272-SFX/10801902195217/33F23B0321, 33F23D0859, 33F23H0500, 33F23K0177, 33F23L0943, 33F24A0280, 33F24C0083, 33F24H0042, 33F24H0260, 33F24J0022, 33F24J0038, 33F24J0088, 33F24L0177, 33F25C0028, 33F25E1090, 33F25F0301, 33F25F0514; CS-15312-SFX/30801902197529/33F23G0541, 33F23H0984, 33F23L0951, 33F24A0283; CS-15422-SFX/30801902195327/33F23K0179, 33F23L0956, 33F24A0290, 33F24E0814; CS-15502-SFX/30801902195563/33F23L0959, 33F24A0286, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2159-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.