On 2012-08-15 the FDA classified a Class II recall of ***REF IFNT611***Full OSSEOTITE Tapered Certain Implant 6 x 11.5mm***Sterile by radiation***Rx Only***BIOMET 3i Dental Iberica S.L. WTC Alm… by Biomet 3i, citing biomet 3i is recalling their product Full Osseotite Tapered Certain Implant IFNT611. The depth of the implant's internal hex is too shallow. The condition, if present, may prevent….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2158-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-03-23 |
| Published | 2012-08-15 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Biomet 3i |
| Distribution | n/a |
Product
***REF IFNT611***Full OSSEOTITE Tapered Certain Implant 6 x 11.5mm***Sterile by radiation***Rx Only***BIOMET 3i Dental Iberica S.L. WTC Almeda Park, Ed, 1, Planta 1 Pl. de la Pau s/n08940 Cornella de Llobregat (Barcelona) Spain. Lot #2011110798. Intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully endentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridge work and to retain over dentures.
Reason
Biomet 3i is recalling their product Full Osseotite Tapered Certain Implant IFNT611. The depth of the implant's internal hex is too shallow. The condition, if present, may prevent the driver; cover screw or abutment from fully engaging the implant.
Identifiers
| code_info | Lot Number 2011110798 |
|---|