FDA Device recall Z-2157-2017

UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322

On 2017-05-31 the FDA classified a Class II recall of UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322 by Beckman Coulter, citing there is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2157-2017
ClassificationClass II
Statusterminated
Initiated2017-04-24
Published2017-05-31
Last updated2026-09-13 11:54 UTC
CompanyBeckman Coulter
DistributionUS

Product

UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.

Reason

There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.

Identifiers

code_infoall software versions